Using ProLink
®
Software with Micro Motion
®
Transmitters
143
Calibrate Menu
continued
Density calibration
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
Figure 14-5. Density Point 2 Calibration dialog box for RFT9739
Table 14-4. Density of water
5. To perform the third-point density calibration, follow these steps:
a. Compare the maximum flow rate allowed by the process against
the appropriate value from
Table 14-1
. If the maximum flow rate
allowed by the process is less than the value listed in
Table 14-1
,
it is not necessary to perform the third-point (K3) calibration.
b. Fill the sensor with a process fluid that has a stable density.
Temperature Density Temperature Density Temperature Density
°
F
°
Cg/cc
°
F
°
Cg/cc
°
F
°
Cg/cc
32
33
34
35
36
37
38
39
0.0
0.6
1.1
1.7
2.2
2.8
3.3
3.9
0.9998
0.9998
0.9999
0.9999
0.9999
0.9999
0.9999
1.0000
56
57
58
59
60
61
62
63
13.3
13.9
14.4
15.0
15.6
16.1
16.7
17.2
0.9994
0.9992
0.9992
0.9991
0.9991
0.9989
0.9989
0.9988
80
81
82
83
84
85
86
95
26.7
27.2
27.8
28.3
28.9
29.4
30
35
0.9966
0.9964
0.9963
0.9961
0.9960
0.9958
0.9956
0.9941
40
41
42
43
44
45
46
47
4.4
5.0
5.6
6.1
6.7
7.2
7.8
8.3
1.0000
0.9999
0.9999
0.9999
0.9999
0.9999
0.9999
0.9998
64
65
66
67
68
69
70
71
17.8
18.3
18.9
19.4
20.0
20.6
21.1
21.7
0.9987
0.9986
0.9984
0.9983
0.9982
0.9981
0.9980
0.9980
100
104
113
122
131
140
149
158
38
40
45
50
55
60
65
70
0.9930
0.9922
0.9902
0.9881
0.9857
0.9832
0.9806
0.9778
48
49
50
51
52
53
54
55
8.9
9.4
10.0
10.6
11.1
11.7
12.2
12.8
0.9998
0.9998
0.9997
0.9996
0.9996
0.9995
0.9995
0.9994
72
73
74
75
76
77
78
79
22.2
22.8
23.3
23.9
24.4
25.0
25.6
26.1
0.9979
0.9977
0.9975
0.9973
0.9972
0.9970
0.9969
0.9968
167
176
185
194
203
212
75
80
85
90
95
100
0.9749
0.9718
0.9686
0.9653
0.9619
0.9584
Komentáře k této Příručce