Micro Motion RFT9709 Manuál s instrukcemi Strana 78

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 214
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 77
64
Using ProLink
®
Software with Micro Motion
®
Transmitters
View Menu: Status
continued
Operational indicators
To program slug flow limits
, follow these steps. For more information,
see
Section 8.5
, page 81.
1. Open the Configure menu, then choose Characterize.
2. In the Characterize Sensor dialog box, at the Low and High text
boxes in the Slug Flow Limits frame, enter the desired slug flow limits
in grams per cubic centimeter (g/cc).
3. Click OK.
In some applications, slug flow typically occurs for short periods of time.
If slug flow ceases in less than one minute, an RFT9739 can continue
holding the last accurately measured flow value until process density
stabilizes within the programmed slug flow limits. A user-defined slug
duration specifies the amount of time the RFT9739 indicates the last
measured flow value before indicating zero flow. (Some RFT9739
software versions cannot indicate last measured value.)
To program a slug duration for an RFT9739
, follow these steps. For
more information,
Section 10.5
, page 116.
1. Open the Configure menu, then choose Transmitter Outputs.
2. Enter the desired time, from 0.00 to 60.00 seconds, into the Slug
Duration text box.
3. Click OK.
Analog and frequency
saturated indicators
If an output variable exceeds its upper range limit, a frequency saturated
or analog saturated indicator switches ON. The indicator can mean the
output variable has exceeded appropriate limits for the process, or the
user needs to change measurement units.
Analog output saturated
If Analog 1 Saturated, Analog 2 Saturated, or Analog Output Saturated
switches ON, open the View menu, then choose Output Levels to open
the Output Levels window.
If the output variable is outside control limits:
use a control device to
decrease the flow rate, density, temperature, or pressure of the fluid.
If the output variable is within control limits:
set the flow loop for
manual operation, then follow one of the procedures described below to
change the span of the milliamp output.
Rescale the milliamp output by adjusting the 4 mA and/or 20 mA
range limit: Open the Configure menu, then choose Transmitter
Outputs. Low Limit, High Limit, and Minimum Span, displayed at the
bottom of the dialog box, indicate sensor limits in the measurement
units established for milliamp output variables.
-
The value entered into the 4 mA text box must be lower than the
value entered into the 20 mA text box.
-
The value entered into the 20 mA text box must be higher than
the displayed low limit and lower than the displayed high limit.
-
The difference between the values entered into the text boxes
must be greater than the displayed minimum span.
Zobrazit stránku 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 213 214

Komentáře k této Příručce

Žádné komentáře