Using ProLink
®
Software with Micro Motion
®
Transmitters
151
Calibrate Menu
continued
Milliamp output trim
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
4. Recalibrate the flowmeter:
a. Perform the flow calibration procedure described in
Section 8.2
,
page 73.
b. Perform the density calibration procedure described in
Section 14.4
, page 138.
c. Perform the viscosity calibration procedure described in
Section 14.5
, page 148.
14.6 Milliamp output trim
Milliamp output trim adjusts the transmitter digital-to-analog converter to
match milliamp outputs with a specific reference standard, receiver, or
readout device.
Preparing for milliamp
output trim
For trimming, milliamp outputs require a reference device such as a
digital multimeter (DMM). To trim the milliamp output, connect the
reference device to the transmitter terminals listed in
Table 14-5
, then
perform the milliamp output trim procedure as described below.
If HART over Bell 202 is being used, and the output being trimmed is the
primary milliamp (mA) output, connect the reference device to the
transmitter as shown in
Figure 14-13
.
Table 14-5. Milliamp output terminals
Transmitter
Primary mA output
terminals
Secondary mA output
terminals
RFT9739 field-mount 17 PV+ 19 SV+
18 PV– 20 SV–
RFT9739 rack-mount CN2-Z30 PV+ CN2-Z28 SV+
CN2-D30 PV– CN2-D28 SV–
IFT9701 4-20 mA + —
4-20 mA –
RFT9712 17 4-20 mA + —
16 4-20 mA –
RFT9729 B16 mA + —
B14 mA –