Using ProLink
®
Software with Micro Motion
®
Transmitters
155
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
15 Test Menu
15.1 Overview
The Test menu enables testing of milliamp outputs at the low output level
(0 or 4 mA), at the high output level (20 mA), or at a specified level from
0 to 22 mA for an RFT9739, or from 2 to 22 mA for an IFT9701,
IFT9703, RFT9712, or RFT9729. A milliamp output test requires a
reference device such as a digital multimeter (DMM).
The Test menu also enables testing of the frequency/pulse output at any
level from 0.1 Hz to 15 kHz for an RFT9739, from 0.01 to 7.2 kHz for an
IFT9701 or IFT9703 version 1.2 or higher, from 0.01 to 1.2 kHz for an
IFT9701 or IFT9703 version lower than 1.2, or from 1 to 10 kHz for an
RFT9712 or RFT9729. A frequency/pulse output test requires a
frequency counter.
The loop test will work only if the transmitter is connected to a sensor.
The loop test will not work if either the transmitter or sensor is in a critical
failure mode.
CAUTION
Failure to set control devices for manual operation
before characterization could cause transmitter
output error.
Before opening the File, Configure, Calibrate, or
Applications menus, set process control devices for
manual operation.
Whenever a dialog box such as the one depicted
below appears, isolate the transmitter from devices
that used transmitter outputs for process control, then
choose Yes.