Using ProLink
®
Software with Micro Motion
®
Transmitters
157
Test Menu
continued
Frequency/pulse output testing
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
Figure 15-1. Test Milliamp Outputs dialog box
If the Bell 202 standard is being used over the primary mA output,
communication failure will occur if output drops below 2 mA. If
communication between the ProLink program and the transmitter is
interrupted before completing the test, one of the following indicators in
the Status window switches ON:
• Analog 1 Fixed
• Analog 2 Fixed
• Analog Output Fixed
To view the Status window, open the View menu, then choose
Transmitter Status. After checking indicators in the Status window,
reopen the Test menu and complete the milliamp output test.
15.3 Frequency/pulse output
testing
Frequency/pulse output testing requires a frequency counter. To test the
output, connect the frequency counter to the transmitter frequency/pulse
output terminals listed in
Table 15-2
, then perform the frequency/pulse
output test procedure.
Table 15-2. Frequency/pulse
output terminals
If communication is lost while in fixed frequency mode, the transmitter
remains in fixed frequency mode. No status indicates this state.
Transmitter Terminals
RFT9739 field-mount 15 Freq +
16 Return
RFT9739 rack-mount CN2-D24 Freq
CN2-D26 Return
IFT9701/IFT9703 +
–
RFT9712 19 Freq +
18 Freq F/R –
RFT9729 CN2-28b Freq +
CN2-24b Freq F/R –