Using ProLink
®
Software with Micro Motion
®
Transmitters
Getting Started
continued
Communication standards
2.2 Communication standards
Switches and jumpers on the transmitter determine the communication
standard used by the transmitter. Micro Motion configures each
transmitter's default communication settings at the factory.
Depending on the transmitter model, transmitters can communicate
using HART and/or Modbus
®
protocol, using the Bell 202 or RS-485
standard. Communication configuration for the ProLink program and
transmitter must be the same.
Factory-default settings
The factory default settings for Version 3 RFT9739 transmitters are:
HART protocol over the Bell 202 standard at 1200 baud, 1 stop bit,
odd parity
Modbus protocol over the RS-485 standard at 9600 baud, 1 stop bit,
odd parity
The factory default settings for the IFT9701, IFT9703, RFT9712, and
RFT9729 are: HART protocol over the Bell 202 standard at 1200 baud, 1
stop bit, odd parity.
User configuration
Transmitter models RFT9739, RFT9712, and RFT9729 can be
reconfigured for user-defined communication settings using switches
and jumpers on the transmitter. To establish a user-defined
communication configuration, see the transmitter instruction manual.
HART
®
and Modbus
®
communication
The primary variable milliamp output on the RFT9739, and the 4-20 mA
output on the IFT9701, IFT9703, RFT9712, and RFT9729 can produce
HART-compatible signals for Bell 202 communication.
Micro Motion transmitters can function as part of a Bell 202 or RS-485
multidrop network.
The RFT9739 can use the Bell 202 or RS-485 standard under HART
protocol, or the RS-485 standard under Modbus protocol.
The IFT9701 and IFT9703 can use the Bell 202 standard under
HART protocol only.
The RFT9712 and RFT9729 can use the Bell 202 or RS-485
standard under HART protocol only.
Up to 15 transmitters can participate with other devices in a Bell 202
multidrop network. Each transmitter must have a unique polling address
of 1 to 15, or a unique tag name.
Up to 32 transmitters can participate in an RS-485 multidrop network.
Each transmitter must have a unique tag name; up to 15 transmitters
may have unique polling addresses from 1 to 15. The IFT9701 and
IFT9703 cannot communicate in an RS-485 network.